CRS-VLB vertical combustion chamber and standards
Test criteria: in accordance with UL62-3839, UL1581-1060-1080, 1100,1140 standard production, testing the flame resistance of the wire.
The instrument consists of: The equipment consists of a control box, vertical combustion chamber, level combustion chamber, burning lamps, solenoid valves, high pressure igniter, gas pipes, pressure regulators, and signal control lines. The control box features: the Spitfire timer Minting timer, burning counter, manual / automatic operation selector switch, vertical / level select switch and power switch.
The main parameters:
Vertical combustion chamber with the level of combustion chamber: Cabinet size 305x355x610mm SUB304 made of stainless steel;
2, the vertical combustion Spitfire tube angle: with the sample angle of 90 degrees;
The level the burning Spitfire tube angle: the sample included angle of 20 degrees;
, Stop: When the data reaches the preset, the machine will automatically stop the test;
4, Annex: flame height ruler of two (500w1 only, 225w1);
The 5 vertical blowtorch: 500w1;
6, standards blowtorch: 225w1;
7, combustion timing :0.01-99 .99S/M/H continuously adjustable;
8 :0.01-99 .99S/M/H stop the clock continuously adjustable;
9, counter :1-9999 can be set;
10, manual and automatic mode selection;
11 Ignition mode: automatic ignition;
12, power supply: AC220V, 50Hz, 5A.